CVE-2026-7251

CRITICAL
Published May 26, 2026 Modified May 26, 2026 CWE-259

Description

Eppendorf BioFlo 320 is vulnerable to due to VNC server using a hard-coded password. If a remote attacker knows the network address of any BioFlo 320 model with remote access enabled, they can gain full control of the user interface by using this password. Once connected, the attacker would have full access to all control panel features for the BioFlo 320. VNC traffic is not encrypted.

CVSS v3.1 Score

9.8
CRITICAL
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H

Weakness Type (CWE)

CWE-259 CWE-259

References

Frequently Asked Questions

What is CVE-2026-7251? +
Eppendorf BioFlo 320 is vulnerable to due to VNC server using a hard-coded password. If a remote attacker knows the network address of any BioFlo 320 model with remote access enabled, they can gain full control of the user interface by using this password. Once connected, the attacker would have full access to all control panel features for the BioFlo 320. VNC traffic is not encrypted. It has a CVSS v3.1 base score of 9.8 (CRITICAL).
How severe is CVE-2026-7251? +
CVE-2026-7251 has a CVSS v3.1 score of 9.8 out of 10, rated CRITICAL. This is a critical vulnerability that should be patched immediately.
How do I check if I'm vulnerable to CVE-2026-7251? +
You can use Secably's free Website Scanner to check your website for known vulnerabilities. For infrastructure scanning, use the Port Scanner to identify exposed services that may be affected. Check the vendor advisories linked above for specific patch and version information.